Diagnostic spirometer with oximetry option
Spirometer
- Records best 3 trials
- Up to 8 blows on one screen
- Internal temperature sensor for BTPS conversion
Finger Oximeter (option)
- Short or long term (overnight) SpO2 and Pulse Rate measurements
- ODI, NOD, T90%, T89%, T88, T87%
- Sleep oximetry with desaturation events
- Oximetry during exercise test
FVC, VC with breathing pattern plus MVV tests with real time curves
- High resolution colour screen
- Fast but silent thermal printer
- Digital turbine flow meter with guaranteed accuracy in all environmental conditions
- PRE-POST bronchodilator comparison
- Selectable language and predicted values
- Connectivity: USB, Bluetooth® and RS232
Spirolab III Spirometer Technical Specifications
|
Power supply:
|
Rechargeable battery and mains power
|
|
Temperature sensor:
|
Semiconductor (0-45°C)
|
|
Flow sensor:
|
Bi-directional digital turbine
|
|
Flow range:
|
± 16 L/s
|
|
Volume accuracy:
|
± 3% or 50 mL
|
|
Flow accuracy:
|
± 5% or 200 mL/s
|
|
Dynamic resistance:
|
< 0.5 cmH2O/L/s
|
|
Connectivity:
|
USB, Bluetooth, RS 232
|
|
Display:
|
FSTN graphic, 320x240 pixel
|
|
Printer/paper:
|
Thermal, 112 mm width
|
|
Mouthpieces:
|
30 mm external diameter
|
|
Dimensions:
|
310 x 205 x 65 mm
|
|
Weight:
|
1,9 Kg cira
|
Measured parameters
FVC, FEV1, FEV1/FVC%, FEV6, FEV1/FEV6%, PEF, FEF25%, FEF50%, FEF75%, FEF25-75%, FET, Vext, FIVC, FIV1, FIV1/FIVC%, PIF, *FVC, *FEV1, *PEF, VC, IVC, IC, ERV, FEV1/VC%, VT, VE, Rf, ti, te, ti/t-tot, VT/ti, MVV
(* Best value)
Optional: Finger oximeter technical specifications
|
SpO2 range:
|
0-99%
|
|
SpO2 accuracy:
|
± 2% between 70-99% SpO2
|
|
Pulse Rate range:
|
30-254 BPM
|
|
Pulse Rate accuracy:
|
± 2 BPM or 2 %
|
winspiroPRO Software
- SpO2 and Pulse Rate graphic trend
- Flexible reporting with several printout categories
- Statistical analysis of desaturation events
|